Available for license.

(Currently under development)

Main performance specification
Flow rate 0 to 100 µl/h This is the linear range, stroke volume is constant and the flow rate is proportional to the frequency (Frequency range 0-0.2Hz).
Priming rate 250 µl/h This is the maximum flow rate in the uncontrolled range.
(Frequency 2Hz).
Stroke volume 150 nl Minimum dosage unit.
Leakage Non measurable
< 0.05 ml/h
Forward and backward leak with 150 mbar overpressure applied on inlet respectively on outlet.
Accuracy ± 5% Overall accuracy of the stroke volume within nominal conditions.
Inlet pressure
Outlet pressure
-350 to +350 mbar
-400 to +200 mbar
 
Viscosity 0-10 mPa.s  
Actuation voltage -40V/ +110V Voltage applied to the piezo disc.
Compression ratio e=deltaV/V0=1.15 The pump is self priming and tolerant to small air bubbles.
Reliability 1.8 108 stroke
67 days
(equivalent to 25yr of use). Accelerated dry test at 10Hz : no measurable degradation.
Real time liquid flow test 0.05Hz (see plot below).
Chip dimensions 16x12x1.86mm  

<< Back




Debiotech S.A
., Av. de Sévelin 28, 1004 Lausanne, Switzerland.
© Copyright Debiotech S.A. all rights reserved.